First Call Parts is dedicated to delivering cooling units of the highest quality. Our comprehensive process ensures that every cooling unit meets the highest performance standards before it reaches you. Here’s a detailed overview of how we uphold our commitment to quality:

5764555 Siemens Cath/Angio Cooling Unit

5764555STD

Standard Connection

5764555QC

Brass Quick Connect

5764555KKA

Stainless Steel Quick Connect

Cooling Unit Refurbishment

Our Quality Assurance Process

Initial Incoming Inspection:

  • Thorough Physical Inspection: We assess wear-and-tear from usage and identify any physical repairs needed.

Rework Process:

  • The system is flushed and flowing clear before parts installed
  • Defective components repaired
  • Replace pump & hoses
  • Replace radiators
  • Production work on covers

Final Checks:

  • Testing: System tests are performed,
  • Final Cleaning and Documentation: The unit’s exterior is cleaned, tamper indicator stickers are applied, and all documentation is updated and archived before transfer to Finished Goods.

**Important Note of Caution Prior to Replacing this Cooling Unit** If you are replacing your cooling unit because the tube oil pump is not turning (no 24VDC at kluver tube chiller pins 1-2)

  • The K2 relay in the system cabinet is known to fail in a way that will chatter 230VAC and damage the Kluver circuit board
  • If this happens the K2 relay must be replaced and it is located in the system cabinet.
  • This relay failure is usually intermittent and may be very difficult to catch in the failed state. Take a look at the diagram, here.

Our meticulous process ensures that every cooling unit we deliver performs at its best and provides the reliability you expect. Should you have any questions or need further details, please do not hesitate to reach out.

 