In the AACE software, a dual calibration can be performed. This allows for samples off-scale to be run on a previously run higher concentration calibration curve. In the run, tick the box on the main page listed as "Dual-cal run". Then select the curve to be used for the higher calibration. Note both calibrations must have the same number of channels in use.
On the channel tab, both calibrations will be shown. In addition, a QC cup (QC10) must be added into the run. This QC10 cup needs to be the same concentration as the Drift peak from the higher calibration. This will normalize the results.
A flag in the report (H) will show that this sample concentration was run on the higher range calibration.
For more information about setting this up in the software send us an
email.