The Care and Feeding of Healthcare Steam Sterilizers
We have all heard the phrase, “You are what you eat.” Proper nutrition is important for a healthy life. Analogously speaking, healthcare steam sterilizers have similar “nutritional” requirements that, if followed, will enable them to provide you with sterile, stain-free, dry loads. This article will give you some guidance on how best to “feed” your sterilizers and get the results you need out of them.
But what are the nutritional requirements? Well, the requirements are water, steam, electricity, and compressed air (in some cases). Before we can discuss what the requirements for each of these are to achieve the best results, let’s look at how each of them is used in the steam sterilization process.
Water
Water is used for three things in a steam sterilizer. One is to cool effluent (hot water and steam) released to the drain or to a recirculation tank. The other is to provide the sealing or motive fluid for the vacuum system. The third is to generate steam if the sterilizer has an on-board boiler. There is also the issue of “utility” and “critical” water. Let’s look at the first two of these uses in more detail.
Cooling
Water that is used for cooling has no real requirements on it except that it be cold enough for the purpose of cooling. “Cold enough for the purpose of cooling” is a bit of a can of worms, since “cold enough” can mean a lot of different temperatures depending on what has to be cooled.
If it is a drain to the floor drain of the technical area behind the sterilizers, it has to be cold enough to cool the effluent to less than the maximum allowed drain temperature by plumbing code, 140°F. Since this is water that you are just throwing down the drain, you would like to use as little as possible for efficiency’s sake. Which means that you want it to be as cold as possible. If your water supply is not very cold, say about 75-80°F, you may want to consider cooling it if water is expensive where you are located and electricity less so. A cost/benefit analysis should be undertaken for this purpose. Note that only utility water should be used for drain cooling.
The same issue arises for the second use of water, as the motive/sealing fluid in the sterilizer’s vacuum system. Again, utility water is the type of water that should be used in this case.
Water can be used to generate vacuum by flowing through a venturi (aka, ejector or eductor) at a high flow rate, which aspirates air from the sidearm of the venturi. This method uses a lot of water but has no need for a vacuum pump, making it mechanically simple to implement. The flow rate and temperature of the water will determine the pumping speed, with higher flow rate and lower temperature improving pumping speed.
Water is also used in steam sterilizers to generate vacuum in a “liquid ring pump.” This type of vacuum pump uses a lot less water, but typically needs three-phase electrical power.
Recirculation systems use a lot less water than once-through systems and can be used with either venturis or liquid-ring pumps. The water can be recirculated to a holding tank whose temperature is controlled by adding cold water to it until its temperature falls below a preset value. Some sterilizer manufacturers offer this as an add-on option. Some build it in. The temperature can also be kept below the preset value by the use of an electrical cooler for the recirculation tank. Again, a total cost analysis should be made before an order is placed.
Steam Generation
Steam in healthcare facilities is typically generated in large, centralized boilers that supply the many steam needs of the hospital. This type of steam is called “plant steam” or “house steam.” The starting point for plant steam is utility water. The chemical needs of the boiler and steam plumbing system may result in steam that is not suitable for sterilization, since it may cause staining or corrosion. This is because of the boiler water additives that are needed to avoid corrosion of the (usually) iron plumbing. Plant steam that is ideal for sterilization has few impurities and a pH from 7.5 to 8.5. The permitted levels of impurities are defined in ANSI/AAMI ST108:2023, Water for processing of medical devices.
If a local boiler is used, and if the sterilizer has stainless steel plumbing and valves, critical water can be used to generate steam that is dedicated to sterilization, supplying only the sterilizers. These can be electrically heated or use “indirect heating”, using house steam to boil the water in the boiler via heat exchanger tubing. This is a much cleaner solution for steam sterilization, but may require more attention and, depending upon the location of the healthcare facility, a dedicated engineer to monitor the boiler whenever it is in use.
This is the first in a series of articles designed to demystify the technology behind cleaning, disinfection and sterilization. I hope that you find these useful and would welcome any questions.
|